Omron Healthcare Blood Pressure Monitor BP742 User Manual

INSTRUCTION MANUAL  
5 series  
Blood Pressure Monitor  
Model BP742  
 
INTRODUCTION  
®
®
Thank you for purchasing the OMRON BP742 IntelliSense Automatic Blood  
Pressure Monitor.  
Fill in for future reference.  
DATE PURCHASED: __________________  
SERIAL NUMBER: ____________________  
• Staple your purchase receipt here  
• Register your product on-line at  
Your new digital blood pressure monitor uses the oscillometric method of blood  
pressure measurement. This means the monitor detects your blood’s movement through  
your brachial artery and converts the movements into a digital reading. An  
oscillometric monitor does not need a stethoscope so the monitor is simple to use.  
The BP742 comes with the following components:  
• Monitor • Standard Adult Arm Cuff • 4 “AA” Alkaline Batteries  
• Instruction Manual printed in English and Spanish • Quick Start Guide  
®
The BP742 IntelliSense Automatic Blood Pressure Monitor is intended for home use.  
SAVE THESE INSTRUCTIONS  
Download from Www.Somanuals.com3. All Manuals Search And Download.  
SAFETY INFORMATION  
To assure the correct use of the product, basic safety measures should always be  
followed including the warnings and cautions listed in this instruction manual.  
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL  
Indicates a potentially hazardous situation which, if not avoided,  
could result in death or serious injury.  
WARNING  
Indicates a potentially hazardous situation which, if not avoided,  
may result in minor or moderate injury to the user or patient or  
damage to the equipment or other property.  
CAUTION  
OPERATING THE DEVICE  
Contact your physician for specific information about your blood pressure.  
Self-diagnosis and treatment using measured results may be dangerous. Follow the  
instructions of your physician or licensed healthcare provider.  
The monitor is not intended to be a diagnostic device.  
If battery fluid should get in your eyes, immediately rinse with plenty of clean  
water. Contact a physician immediately.  
Read all of the information in the instruction manual and any other literature in the  
box before operating the unit.  
This device is intended for use in measuring blood pressure and pulse rate in the  
adult population. Do not use this device on infants or persons who cannot express  
their intentions.  
Read the Special Conditions section of the instruction manual if your systolic  
pressure is known to be more than 220 mmHg. Inflating to a higher pressure than  
necessary may result in bruising where the cuff is applied.  
Operate the device only as intended. Do not use the device for any other purpose.  
Dispose of the device, components and optional accessories according to applicable  
local regulations. Unlawful disposal may cause environmental pollution.  
 
4
SAFETY INFORMATION  
Do not use a cellular phone near the device. It may result in an operational failure.  
Use only Omron authorized parts and accessories. Parts and accessories not  
approved for use with the device may damage the unit.  
Use only 1.5 V “AA” alkaline batteries with this device. Do not use other types of  
batteries. This may damage the unit  
CARE AND MAINTENANCE  
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.  
Do not submerge the device or any of the components in water.  
Store the device and the components in a clean, safe location.  
Changes or modification not approved by the manufacturer will void the user  
warranty. Do not disassemble or attempt to repair the unit or components.  
Download from Www.Somanuals.com5. All Manuals Search And Download.  
BEFORE TAKING A MEASUREMENT  
To ensure a reliable reading follow these recommendations:  
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes  
before taking a measurement. Rest for at least 15 minutes before taking the  
measurement.  
2. Stress raises blood pressure. Avoid taking measurements during stressful times.  
3. Measurements should be taken in a quiet place.  
4. Remove tight-fitting clothing from your left arm.  
5. Sit on a chair with your feet flat on the floor. Rest your left arm on table so that the  
cuff is at the same level as your heart.  
6. Remain still and do not talk during the measurement.  
7. Wait 2-3 minutes between measurements. The wait time allows the arteries to return  
to the condition prior to taking the blood pressure measurement. You may need to  
increase the wait time depending on your individual physiological characteristics.  
8. Keep a record of your blood pressure and pulse readings for your physician. A  
single measurement does not provide an accurate indication of your true blood  
pressure. You need to take and record several readings over a period of time. Try to  
measure your blood pressure at the same time each day for consistency.  
 
6
KNOW YOUR UNIT  
Main Unit:  
Display  
Battery compartment  
START/STOP button  
Air Jack  
USER ID  
Selection Switch  
AC adapter Jack  
Up/Down (  
) buttons  
Date/Time setting (  
button  
)
Components:  
Automatic  
Blood Pressure Monitor  
5
series  
Air tube  
Air plug  
Model BP742  
5
Series Arm Blood Pressure  
Monitor  
BP742  
Instruction  
Manual  
Quick Start  
Guide  
Standard Adult Arm Cuff  
UNIT DISPLAY  
Systolic blood pressure  
Diastolic blood pressure  
Heartbeat symbol /  
Memory symbol  
(Displayed when viewing  
values stored in memory)  
Movement error symbol  
Hypertension Indicator  
Pulse display  
Deflation symbol  
User ID symbol (A/B)  
Irregular Heartbeat symbol  
Date/Time Display  
Battery low indicator  
Download from Www.Somanuals.com7. All Manuals Search And Download.  
DISPLAY SYMBOLS  
IRREGULAR HEARTBEAT SYMBOL (  
)
When the monitor detects an irregular rhythm two or more times during the  
measurement, the Irregular Heartbeat Symbol (  
measurement values.  
) will appear on the display with the  
An irregular heartbeat rhythm is defined  
Normal Heartbeat  
as a rhythm that is more than 25%  
Pulse  
slower or 25% faster from the average  
rhythm detected while the monitor is  
Blood  
pressure  
measuring the systolic and diastolic  
blood pressure.  
Irregular Heartbeat  
If the irregular heartbeat symbol (  
displays with your measurement results,  
)
Short Long  
Pulse  
we recommend you consult your  
Blood  
pressure  
physician. Follow the directions of your  
doctor.  
MOVEMENT ERROR SYMBOL (  
)
The Movement Error Symbol (  
) is displayed if you move your body during the  
measurement. Please remove the arm cuff, and wait 2-3 minutes. Reapply the arm cuff  
and take another measurement.  
HEARTBEAT SYMBOL/HYPERTENSION INDICATOR (  
TAKING A MEASUREMENT  
)
The Heartbeat Symbol ( ) flashes on the display during the measurement. The  
Heartbeat Symbol ( ) flashes at every heartbeat.  
When the measurement is complete, the Heartbeat Symbol ( ) flashes on the  
display with your blood pressure and pulse rate if the reading is above 135 for the  
Systolic Blood Pressure value and/or above 85 for the Diastolic Blood Pressure  
value.  
USING THE MEMORY FUNCTION  
The Heartbeat Symbol ( ) flashes when the blood pressure reading stored in the  
memory is above 135 for the Systolic Blood Pressure value and/or above 85 for the  
Diastolic Blood Pressure value.  
 
8
DISPLAY SYMBOLS  
The American Heart Association recommends the following guideline for upper limit  
of normal for home blood pressure readings:  
AHA Home Guideline for Upper Limit of Normal BP  
Systolic Blood Pressure  
Diastolic Blood Pressure  
135 mmHg  
85 mmHg  
This is a general guideline because blood pressure depends on age and morbidity.  
The AHA recommends a lower home blood pressure goal for certain patients,  
1
including diabetic patients, pregnant women, and patients with renal failure .  
WARNING  
The monitor is not intended to be a diagnostic device.  
WARNING  
Contact your physician for specific information about your blood pressure.  
Self-diagnosis and treatment using measured results may be dangerous.  
Follow the instructions of your physician or licensed healthcare provider.  
Omron Healthcare recommends you contact your physician if the Heartbeat Symbol
(
) flashes on the display with your measurement values.  
1American Heart Association Scientific Statement 2004  
Download from Www.Somanuals.com9. All Manuals Search And Download.  
BATTERY INSTALLATION  
CAUTION  
Use only 1.5V “AA” alkaline batteries with this device.  
1. Press the indicator on the battery cover and �  
slide the cover off in the direction of the arrow.  
2. Install 4 “AA” size batteries so the + (positive) and
- (negative) polarities match the polarities of �  
the battery compartment as indicated.  
3. Replace the battery cover.  
 
10  
BATTERY INSTALLATION  
BATTERY REPLACEMENT  
Low Battery Indicator  
When the Low Battery Indicator appears on the display, turn the monitor off and  
remove all the batteries. Replace with four new batteries at the same time. Long-life  
alkaline batteries are recommended.  
WARNING  
If battery fluid should get in your eyes, immediately rinse with  
plenty of clean water. Contact a physician immediately.  
NOTES: • Turn the unit off before replacing the batteries. If the batteries are removed  
while the unit is still on, the date and time will be reset to that of the  
previous use. However, the measurement results are not deleted.  
• When the batteries are replaced, you may need to reset the date and time. If  
“-:--” appears on the display, refer to “Setting the Date and Time”.  
• The supplied batteries may have a shorter life.  
CAUTION  
Dispose of the device, components and optional accessories according to  
applicable local regulations. Unlawful disposal may cause environmental  
pollution.  
Download from Www.Somanuals.co1m1. All Manuals Search And Download.  
SETTING THE DATE AND TIME  
Set the monitor to the current date and time before taking a measurement for the first  
time. If the date and time has not been set, the measurement values are stored without  
any date or time information.  
1. TO START THE SETTING  
Press the Date/Time Setting button ( ) to adjust the date and time.  
The year flashes on the display  
.
2. SETTING THE YEAR  
The year can be set between 2010 and 2040. When the  
display reaches 2040, it will return to 2010.  
Press the Up button ( ) to advance by increments of one  
year.  
Press the Date/Time Setting button ( ) to set the current  
year.  
The month flashes on the display.  
NOTES: • Press the Down button ( ) to decrease the  
date and time values by increments of one.  
• Press and hold the button to increase or  
decrease the date and time values faster.  
 
12  
SETTING THE DATE AND TIME  
3. SETTING THE MONTH  
Press the Up button ( ) to advance by increments of one month.  
Press the Date/Time Setting button ( ) to set the current  
month.  
The day flashes on the display.  
4. SETTING THE DAY  
Press the Up button ( ) to advance by increments of  
one day.  
Press the Date/Time Setting button ( ) to set the current  
day.  
The hour flashes on the display.  
5. SETTING THE HOUR  
The time is set using AM or PM.  
Press the Up button ( ) to advance by increments of  
one hour.  
Press the Date/Time Setting button ( ) to set the current  
hour.  
The minute flashes on the display.  
6. SETTING THE MINUTE  
Press the Up button ( ) to advance by increments of one  
minute.  
Press the Date/Time Setting button ( ) to set the minute setting.  
7. TO TURN THE MONITOR OFF  
Press the START/STOP button.  
NOTE: The monitor will automatically turn off five minutes after the minute has  
been set.  
Download from Www.Somanuals.co1m3. All Manuals Search And Download.  
APPLYING THE ARM CUFF  
CAUTION  
Read all of the information in the instruction manual and any other  
literature in the box before operating the unit.  
CAUTION  
This device is intended for use in measuring blood pressure and  
pulse rate in the adult population. Do not use this device on infants  
or persons who cannot express their intentions.  
APPLYING THE CUFF ON THE LEFT ARM  
1. Make sure the air plug is securely inserted  
in the main unit.  
2. Remove tight-fitting clothing from your upper  
arm.  
 
14  
APPLYING THE ARM CUFF  
3. Apply the cuff to your left upper arm so
Left Arm  
1/2 inch (1 - 2 cm)  
the arrow is centered on the inside of your arm
and aligned with your middle finger.
The air tube runs down the inside of your arm.
The bottom of the cuff should be
Arrow  
Air Tube  
approximately 1/2 “ above your elbow.  
4. Wrap the cuff firmly in place around your 
arm using the cloth fastening.  
APPLYING THE CUFF ON THE RIGHT ARM  
When taking a measurement using the right arm, use this instruction for Step 3.  
Right Arm  
Apply the cuff to your right upper arm so the  
arrow and the air tube runs along the inside  
of your arm. The bottom of the cuff should be
approximately 1/2 “ above your elbow.  
Arrow  
NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air  
to the cuff.  
Download from Www.Somanuals.co1m5. All Manuals Search And Download.  
TAKING A MEASUREMENT  
1. Press the START/STOP button to turn the monitor on.  
All symbols appear on the display.  
2. Select your USER ID (A or B).  
3. Sit on a chair with your feet flat on the floor.  
Place your arm on a table so the cuff is level  
with your heart.  
Keep still and do not talk during measurement.  
4. Press the START/STOP button.  
The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically  
determines your ideal inflation level.  
Remain still and do not move your arm until the entire measurement process is completed.  
NOTE: To stop the inflation or measurement, push the START/STOP button. The  
monitor will stop inflating, start deflating, and will turn off.  
 
16  
TAKING A MEASUREMENT  
5. Inflation stops automatically and the measurement is started.  
As the cuff deflates, decreasing numbers appear on the display and the Heartbeat  
Symbol ( ) flashes.  
6. When the measurement is complete, the arm cuff completely  
deflates. Your blood pressure and pulse rate are displayed.  
7. Press the START/STOP button to turn the monitor off.  
NOTE: The monitor will automatically turn off after  
five minutes.  
Download from Www.Somanuals.co1m7. All Manuals Search And Download.  
TAKING A MEASUREMENT  
SPECIAL CONDITIONS  
CAUTION  
Inflating to a higher pressure than necessary may result in bruising where  
the cuff is applied.  
If your systolic pressure is known to be more than 220 mmHg, press and  
hold the START/STOP button until the monitor inflates 30 to 40 mmHg  
higher than your expected systolic pressure.  
NOTES: • Do not apply more pressure than necessary.  
• The monitor will not inflate above 299 mmHg.  
When the measurement is complete, the arm cuff completely deflates.  
Your blood pressure and pulse rate are displayed.  
 
18  
USING THE MEMORY FUNCTION  
The monitor is designed to store the blood pressure and the pulse rate in the memory  
for two people (USER A and USER B) every time a measurement is completed.  
The monitor automatically stores up to 30 sets of measurement values (blood pressure  
and pulse rate) for each user (A and B). When 30 sets of measurement values are  
stored, the oldest record is deleted to save the most recent values.  
TO DISPLAY THE MEASUREMENT VALUES  
1. Press the START/STOP button to turn the monitor on.  
2. Select your USER ID (A or B).  
3. Press the Up button ( ) to display the most recent set of measurement values on  
the screen.
Press the Down button ( ) to display the oldest set of measurement values.  
Download from Www.Somanuals.co1m9. All Manuals Search And Download.  
USING THE MEMORY FUNCTION  
NOTES: • The measurement value number is displayed  
briefly in the Pulse Display area when you press  
the Up/Down buttons (  
).  
• The date and time are alternately displayed with  
the measurement values.  
• If the date and time are not set, “-:--” is displayed  
with the measurement values.  
• Press the button repeatedly to display the next  
values.  
• Press and hold the button to display the values  
faster.  
Alternating  
display  
TO DELETE ALL VALUES STORED IN THE MEMORY  
The values stored in the memory are deleted by USER ID.  
You cannot partially delete values stored in the memory. All values for  
the user you select will be deleted.  
1. Press the START/STOP button to turn the monitor on.  
2. Select your USER ID (A or B).  
3. While holding the Down button ( ), press the START/STOP button down for  
more than five seconds until the display appears as it does in the following  
illustration.  
4. Press the START/STOP button to turn the monitor off.  
NOTE: The monitor will automatically turn off after five minutes.  
 
20  
CARE AND MAINTENANCE  
To keep your digital blood pressure monitor in the best condition and protect the unit  
from damage, follow the directions listed below:  
Do not forcefully bend the arm cuff or air tube. Do not fold tightly.  
Clean the monitor with a soft dry cloth.  
Do not use any abrasive or volatile cleaners.  
Do not attempt to clean the cuff.  
CAUTION  
Do not submerge the device or any of the components in water. Do  
not subject the monitor to extreme hot or cold temperatures,  
humidity or direct sunlight.  
CAUTION  
Store the device and the components in a clean, safe location.  
CAUTION  
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.  
Remove the batteries if the unit will not be used for three months or longer. Always  
replace all the batteries with new ones at the same time.  
Use the unit consistent with the instruction provided in this manual.  
Use only authorized parts and accessories.  
Use the unit in a manner consistent with the instructions provided in this manual.  
CAUTION  
Use only Omron authorized parts and accessories. Parts and accessories  
not approved for use with the device may damage the unit.  
CAUTION  
Changes or modification not approved by the manufacturer will void the user  
warranty. Do not disassemble or attempt to repair the unit or components.  
Download from Www.Somanuals.co2m1. All Manuals Search And Download.  
ERROR INDICATORS  
SYMBOL  
CAUSE  
CORRECTION  
Take another measurement and  
remain still until the measurement  
is complete.  
Monitor could not  
detect pulse wave.  
Cuff under-inflated.  
Cuff not applied  
correctly.  
Remove the arm cuff.  
Read “Taking a Measurement”.  
Wait 2-3 minutes.  
Cuff over-inflated.  
Device error.  
Take another measurement.  
Contact Customer Service.  
Movement during �  
measurement.  
Keep still and take another  
measurement.  
Remove the arm cuff. Wait 2-3  
minutes and then take another  
measurement. Repeat the steps in  
section “Applying the Arm Cuff.”  
If this error continues to appear,  
contact your doctor.  
Irregular or weak
pulses are detected.  
Replace the four batteries.  
Refer to “Battery Installation.”  
Batteries are exhausted.  
 
22  
TROUBLESHOOTING TIPS  
PROBLEM  
CAUSES AND SOLUTIONS  
Replace all four batteries with new ones.  
Check the battery installation for proper placement  
of the battery polarities.  
No power.  
No display appears on the unit.  
Blood pressure varies constantly. Many factors  
including stress, time of day, how you wrap the  
cuff, may affect your blood pressure.
Measurement values appear too  
high or too low.  
Review the sections “Before Taking a  
Measurement” and “Taking a Measurement”.  
Download from Www.Somanuals.co2m3. All Manuals Search And Download.  
FCC STATEMENT  
Note:  
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)  
This product has been tested and found to comply with the limits for a Class B digital  
device, pursuant to part 15 of the FCC rules.  
These limits are designed to provide reasonable protection against harmful interference  
in a residential installation. The product generates, uses, and can radiate radio  
frequency energy and, if not installed and used in accordance with the instructions,  
may cause harmful interference to radio communications. However, there is no  
guarantee that interference will not occur in a particular installation. If the product does  
cause harmful interference to radio or television reception, which can be determined by  
turning the product on and off, the user is encouraged to try to correct the interference  
by one or more of the following measures:  
• Reorient or relocate the receiving antenna.  
• Increase the separation between the product and the receiver.  
• Connect the product into an outlet on a circuit different from that to which the  
receiver is connected.  
• Consult the dealer or an experienced radio/TV technician for help.  
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)  
This digital apparatus does not exceed the Class B limits for radio noise emissions  
from digital apparatus as set out in the interference-causing equipment standard entitled  
“Digital Apparatus”, ICES-003 of the Canadian Department of Communications.  
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux  
appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur:  
“Appareils Numériques”, ICES-003 édictée par le minister des communications.  
Changes or modifications not expressly approved by the party responsible for  
compliance could void the user’s authority to operate the equipment.  
 
24  
LIMITED WARRANTIES  
®
Y
our BP742 IntelliSense Automatic Blood Pressure Monitor, excluding the arm cuff, is  
warranted to be free from defects in materials and workmanship appearing within 2 years from  
the date of purchase, when used in accordance with the instructions provided with the monitor.  
The arm cuff is warranted to be free from defects in materials and workmanship appearing  
within one year from the date of purchase when the monitor is used in accordance with the  
instructions provided with the monitor. The above warranties extend only to the original retail  
purchaser.  
We will, at our option, repair or replace without charge any monitor or arm cuff covered by the  
above warranties. Repair or replacement is our only responsibility and your only remedy under  
the above warranties.  
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the address  
of the repair location and the return shipping and handling fee.  
Enclose the Proof of Purchase. Include a letter, with your name, address, phone number, and  
description of the specific problem. Pack the product carefully to prevent damage in transit.  
Because of possible loss in transit, we recommend insuring the product with return receipt  
requested.  
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN  
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANY  
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED  
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR  
PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE  
IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF  
THE ABOVE EXPRESS WARRANTY.  
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL,  
INCIDENTAL, CONSEQUENTAIL OR INDIRECT COSTS, EXPENSES OR  
DAMAGES.  
This warranty provides you with specific legal rights, and you may have other rights that vary by  
jurisdiction. Because of special local requirements, some of the above limitations and exclusions  
may not apply to you.  
FOR CUSTOMER SERVICE  
Visit our web site at:  
Call toll free:  
1-800-634-4350  
Download from Www.Somanuals.co2m5. All Manuals Search And Download.  
SPECIFICATIONS  
Model:  
Display:  
Measurement Range:  
Accuracy/Calibration:  
Inflation:  
Deflation:  
Rapid Air Release:  
Measurement Method:  
Power Source:  
Battery Life:  
Operating Temperature  
/Humidity:  
Storage Temperature  
/Humidity/Air Pressure:  
Main Unit Weight:  
Main Unit Dimensions:  
Cuff Size:  
Cuff Circumference:  
Memory:  
Contents:  
 
26  
SPECIFICATIONS  
Optional Accessories:  
NOTE: These specifications are subject to change without notice.  
Download from Www.Somanuals.co2m7. All Manuals Search And Download.  
Omron Healthcare, Inc.  
1200 Lakeside Drive  
Bannockburn, IL 60015  
www.omronhealthcare.com  
Copyright © 2010 Omron Healthcare, Inc.  
Made in China  
5329417-5A  
 

Miele Clothes Dryer T 8012 C User Manual
Network Technologies Network Cables ST C5VA WL500 User Manual
Nikon Webcam SB7I01B1 User Manual
Nilfisk ALTO Pressure Washer 30CA COMPACT I User Manual
Nintendo Video Game Console DS Lite User Manual
Nortel Networks Water System 214393 A User Manual
Novatel Wireless Network Router MC551 NVTL User Manual
Numark Industries DJ Equipment EM460 User Manual
Optoma Technology Projection Television HDBV5100P User Manual
Panasonic DVD Recorder DMR E100H User Manual