R
INSTRUCTION MANUAL
Automatic Digital Blood Pressure Monitor
with IntelliSense™
Model
HEM-757
DIGITAL BLOOD PRESSURE MONITOR
WITH FUZZY LOGIC HEM-757
SYSTOLIC
DIASTOLIC
READY TO
MEASURE
REPLACE
PULSE
/min
BATTERIES
PROGRESS
ON/OFF
START
INTRODUCTION
Thank you for purchasing the Omron HEM-757 IntelliSenseTM
Automatic Blood Pressure Monitor. Although your blood pressure
fluctuates
throughout the day, no self-adjustment is necessary as the monitor
intelligently adjusts to your changing conditions. The IntelliSenseTM
monitor uses fuzzy logic intelligence to sense both your systolic
and diastolic blood pressure values, and the result is personalized
inflation with every reading. Arm discomfort and misreading due to
improper cuff inflation are virtually eliminated.
Your new OMRON monitoruses the oscillometric method of blood
pressure measurement. This means the monitor detects your
blood's movement through your brachial artery and converts the
movements into a digital reading. At the push of a button, the
monitor automatically fills the cuff with air and displays your
systolic, diastolic, and pulse measurements on an extra large
display panel.
Please read this instruction manual thoroughly before using
the Omron IntelliSenseTM Automatic Blood Pressure Monitor.
For specific information on your own blood pressure,
CONTACT YOUR PHYSICIAN.
1
KNOW YOUR UNIT
IntelliSense™ Progress Indicator
Systolic Blood
Pressure Display
ON/OFF Button
READY TO
DIGITAL BLOOD PRESSURE MONITOR
WITH FUZZY LOGIC HEM-757
ON/OFF
MEASURE
REPLACE
START Button
BATTERIES
PROGRESS
SYSTOLIC
START
DIASTOLIC
PULSE
/min
Memory Call Button
Pulse Display
Air Jack
Diastolic Blood
Pressure Display
Air Plug
(Connects to the air jack)
Arm Cuff
Air Tube
Green Cuff Marker
Carrying Case
2
QUICK REFERENCE GUIDE
1. Avoid eating, smoking, and exercising for at least 30 minutes
before taking a measurement.
2. Put your left arm through the cuff loop making sure that the
bottom edge of the cuff is approximately one-half inch above the
elbow.
3. Pull the end of the cuff so that the entire cuff is evenly tightened
around your arm and press the sewn hook material firmly against
the pile side of the cuff. Make sure that your skin will not be
pinched in the D-ring once you inflate the cuff bladder.
Immediately deflate the cuff bladder and readjust the cuff if
skin becomes pinched.
4. Sit in a chair with your feet flat on the floor and place your left
arm on a table so that the cuff is at the same level as your heart.
5. Press the ON/OFF button.
6. After the Heart Symbol ( ) appears next to a zero on the digital
panel, press the START button.
7. Remain still throughout the entire reading.
8. When measurement is complete, the monitor alternately displays
your blood pressure and pulse on the digital panel and
automatically deflates the cuff.
9. Wait 2-3 minutes before taking another blood pressure
measurement. You may require more rest time between readings
depending on your individual physiological characteristics.
3
BATTERY INSTALLATION/REPLACEMENT
1. Slide the battery cover off in
the direction of the arrow.
2. Install or replace 4 "AA" size
batteries so + (positive) and
– (negative) polarities match
the polarities of the battery
compartment as indicated.
3. Replace the battery cover.
4. If the Low Battery Indicator
(
) appears on the display,
Low Battery Indication
replace all four batteries.
Long-life alkaline batteries
are recommended.
5. Remove the batteries if the
monitor will not be used for
an extended period of time.
4
SUGGESTIONS BEFORE BLOOD PRESSURE MEASUREMENT
1. Avoid eating, smoking, and exercising for at least 30 minutes
before taking a measurement. Also rest for at least 5 minutes
before taking a reading.
2. Stress raises blood pressure. Avoid taking measurements during
stressful times.
3. Remove tight-fitting clothing from your left upper arm.
4. Measurement should be taken
in a quiet place and you
should be in a relaxed, seated
position. Rest your left arm on
the arm of a chair or on a table
so that the cuff is at the same
level as your heart.
5. Remain still and do not talk during the measurement.
6. Keep a record of your blood pressure and pulse measurements
for your doctor. Remember, a single measurement does not
provide an accurate indication of your true blood pressure. You
need to take and record several measurements over a period of
time. Try to measure your blood pressure at the same time each
day for consistency. Blood pressure measurements fluctuate
considerably.
7. Wait 2-3 minutes between successive measurements. Waiting
allows the engorged blood vessels to return to normal. You may
require more rest time between readings depending on your
individual physiological characteristics.
5
HOW TO APPLY THE ARM CUFF
1. Insert the air plug of the arm cuff
into the air jack of the monitor.
2. The cuff should be assembled
correctly when it is removed from
the box. If it is not, pass the end
of the cuff furthest from the
tubing through the metal D-ring
to form a loop. The smooth cloth
should be on the inside of the
Pile Side Material
Sewn Hook Material
cuff loop.
3. If the cuff is assembled correctly,
the sewn hook material material
will be on the outside of the cuff
loop and the metal D-ring will not
touch your skin.
D-Ring
4. Put your left arm through
the cuff loop. The bottom
of the cuff should be
approximately 1/2"
Bottom Edge 1/2 Inch
above the elbow. The
cuff tubing should be
positioned along the
middle of the inside of
your arm as shown.
Cuff Tube
Cuff Tab
5. Pull the cuff so that the top and
bottom edges are tightened
evenly around your arm.
6
HOW TO APPLY THE ARM CUFF
6. When the cuff is positioned
(Left arm)
correctly, press the sewn
hook material FIRMLY
against the pile side of the
cuff.
7. The cuff should be
wrapped so that it fits
snugly and stays in place.
You should be able to fit
one finger between the cuff
and your arm.
8. Sit in a chair with your feet
flat on the floor and rest
your arm on a table so that
the cuff is at the same level
as your heart.
9. Relax your arm and turn
your palm upward.
10. Be sure there are no kinks
in the air tubing.
7
HOW TO TAKE A READING
Please read "A Few Suggestions Before Blood Pressure Measurement" and
"How To Apply The Arm Cuff" before taking a reading.
1. Press the ON/OFF Button.
a) All display symbols appear for
approximately one second. This
is the initial LCD test.
b) The display symbols disappear
and the Deflation Indication
Initial LCD Test
Symbol ( ) starts to flash.
c) When the monitor completes the
necessary preparations before
measurement, the Heart Symbol
(
) appears next to a zero.
Deflation Indication
NOTE: Wait for the Heart Symbol (
before taking a measurement.
)
2. Press the START button. As the
cuff begins to inflate, the
IntelliSense™ monitor
Ready to Measure
automatically determines your
ideal inflation level. During this
process, the IntelliSense™
symbol ( ) will be flashing on
the display. Because this monitor
detects the pulse even during
inflation, do not move your arm
and remain still during the
measurement.
START
Display of measurement process
A
mark indicates the
Inflation started.
measurement process.
When the Start Button is pushed,
a
mark flushes.
8
HOW TO TAKE A READING
3. The IntelliSense™ Progress
Indicator ( ) will flash on the
display showing that the
monitor’s fuzzy logic intelligence
is determining your personal
measurements.
NOTE: Because blood pressure is being
measured even during deflation, it is
important to remain still until the reading
is fully completed.
Display of measurement process
A
mark indicates the
measurement process. During
the inflation, a mark turns on.
During the deflation, a mark
turns on and one each mark
turns off with the progress of
approximately every 1/3 of the
process.
4. When the measurement is
complete, a Heart Symbol (
)
flashes and the arm cuff deflates.
5. At the end of the measurement,
your blood pressure and pulse
are alternately displayed.
6. Press the ON/OFF button to turn
the monitor off.
NOTE: If you forget to turn the monitor
off, the monitor will automatically power
off after five minutes.
ON/OFF
9
HOW TO TAKE A READING
Special instructions regarding the inflation of
the IntelliSense™ Monitor:
1. In rare circumstances the user’s
physiological characteristics may
require a higher level of inflation.
These circumstances may
include weak pulse or an
unusually high systolic blood
pressure. When this occurs, the
monitor may reinflate to a value
of about 30 mmHg higher than
the previous inflation value.
Reinflation occurs only once.
2. If your systolic pressure is known
to be more than 220 mmHg,
push and hold the START button
START
until the monitor inflates 30 to 40
mmHg higher than your
suspected systolic pressure.
Continue to press button
NOTE: Do not apply more
pressure than
necessary.
3. If you want to stop the
measurement, push the ON/OFF
button. The monitor will stop
inflating and start deflating
rapidly, then the monitor will
power off.
10
HOW TO USE THE MEMORY FUNCTION
How to Use the Memory Function
This monitor has the function to store fourteen measurement values.
Push the
Button while a ( ) mark is displayed.
• Maximum of fourteen stored measurement values can be called
out.
• For the fifteenth measurements and after, the values from the
oldest measurement will be deleted.
Previous value
Value before
the previous
measurement
Value of
fourteen
measurements
before
If you want to delete all the stored data
You can delete all the stored data while the ( ) mark is displayed.
START
Push these two buttons at
the same time.
Display of memory deletion
Download from Www.Somanuals.co1m1. All Manuals Search And Download.
HOW TO USE THE OPTIONAL AC ADAPTER
(ACCESSORY ITEM SOLD SEPARATELY)
1. When using AC power, use only the exclusive AC adapter
(Model HEM-ADPT1) which can be purchased from:
Omron Healthcare, Inc.
300 Lakeview Parkway
Vernon Hills, IL 60061
Toll Free Phone #: 1-800-634-4350
Plug
2. Insert the AC adapter plug into
the jack on the side of the
monitor.
3. Insert the AC adapter plug into
a 120V AC outlet (50 - 60
Jack
cycles).
Option AC Adapter
4. To remove the AC adapter, disconnect the adapter plug from the
AC outlet first and then disconnect the cord from the monitor’s
jack. To avoid possible damage to the monitor, use only the
exclusive AC adapter (Model HEM-ADPT1) available from
Omron Healthcare, Inc. Other adapters may vary in output
voltage and polarities.
AC Adapter Output
6 V, 500 mA
NOTE: The monitor is designed not to draw power from the
batteries when the AC adapter is in use.
12
TROUBLESHOOTING
Error Indicator
Cause
Correction
Wait 10 minutes before taking
another measurement. Repeat
steps listed under "How to Take a
Reading." If "EE" is displayed again,
take another measurement by
pressing and holding the START
button as listed in step 2 under
"How to Take a Reading."
Cuff under-inflated.
Movement during
measurement.
Carefully read and repeat steps
listed under "How to Take a
Reading."
Cuff over-inflated (more than
300 mmHg).
Battery voltage is excessively Replace all four "AA" batteries with
low.
new batteries.
Please contact the store where you
have purchased this unit or the
nearest OMRON dealer.
Trouble caused by abnormal
memory function.
If you continue to have difficulty obtaining a reading, call
OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1-800-634-4350.
Please tell the representative you have an OMRON Model HEM-757.
Download from Www.Somanuals.co1m3. All Manuals Search And Download.
CAUTION
CAUTION:
Changes or modifications not expressly approved by Omron Healthcare, Inc.
could void the user’s authority to operate this product.
NOTE:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. The product generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If the product does cause harmful interference to radio or television
reception, which can be determined by turning the product on and off, the user
is encouraged to try to correct the interference by one or more of the following
measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and receiver.
• Connect the product into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise
emissions from digital apparatus as set out in the interference-causing
equipment standard entitled "Digital Apparatus", ICES-003 of the Canadian
Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques
applicables aux appareils numériques de Classe B prescrites dans la norme
sur le matériel brouilleur: "Appareils Numériques", NMB-003 èdictée par le
ministre des communications.
14
CARE AND MAINTENANCE
To protect your monitor from damage, please AVOID the
following:
• Subjecting your monitor to extreme temperatures, humidity,
and direct sunlight.
• Folding the cuff and tubing tightly.
• Inflating the monitor over 280 mmHg.
• Disassembling the monitor.
• Subjecting the monitor to strong shocks (for example,
dropping the monitor on the floor.)
• Do not clean the monitor with volatile liquids.
THE MONITOR SHOULD BE CLEANED WITH A SOFT,
DRY CLOTH.
Download from Www.Somanuals.co1m5. All Manuals Search And Download.
FIVE YEAR LIMITED WARRANTY
Your HEM-757 IntelliSense™ Automatic Blood Pressure Monitor is
warranted to be free from manufacturing defects for a period of five
years under normal use. The five year warranty excludes the
monitor cuff. The cuff is warranted for a one year period. This
warranty extends only to the original retail purchaser.
Should repair be needed within the warranty period, ship the unit
prepaid to Omron Healthcare, Inc., 300 Lakeview Parkway,
Vernon Hills, IL 60061, Attn: Service Dept., together with $5.00
for return shipping and insurance. Be sure to include the model
number of your unit and your phone number on any
correspondence.
We will either repair or replace (at our option) free of charge any
parts necessary to correct defects in the materials or workmanship.
The above warranty is complete and exclusive. The warrantor
expressly disclaims liability for incidental, special, or consequential
damages of any nature. (Some states do not allow the exclusion or
limitation of incidental or consequential damages, so the above
warranty may not apply to you.)
Any implied warranties arising by the operation of law shall be
limited in duration to the term of this warranty. (Some states do not
allow limitations on how long an implied warranty lasts, so the
above limitation may not apply to you.)
This warranty gives you specific legal rights, and you may have
other rights which vary from state to state. As a condition to
operation of your warranty, the enclosed registration card must be
completed and sent to us within 10 days from the date of purchase.
FOR CUSTOMER SERVICE CALL TOLL FREE:
1-800-634-4350
16
SPECIFICATIONS
Model:
HEM-757
Display:
LCD Digital Display
Measurement Range:
Pressure: 0 to 280 mmHg
Pulse: 40 to 180/minute
Pressure: ±3 mmHg or 2% of reading
Pulse: ±5% of reading
Automatic by electric pump
Automatic deflation system
Accuracy/Calibration:
Inflation:
Deflation:
Rapid Pressure Release:
Pressure Detection:
Measurement Method:
Pulse Wave Detection:
Power Source:
Active electromagnetic control valve
Electrostatic capacity type pressure sensor
Oscillometric method
Capacitive pressure sensor
4 "AA" batteries or AC adapter (sold separately)
Approximately 300 uses
Battery Life:
Operating Temperature/Humidity: 50°F to 104°F (10°C to 40°C)
30 to 85% RH maximum
Storage Temperature/Humidity:
–4°F to 140°F (–20°C to 60°C)
10 to 95% RH maximum
Console Weight:
Approximately 19 oz (530 g) not including batteries
Approximately 4 1/2" (l) x 7" (w) x 2 5/6" (h)
(115 mm x 177 mm x 72 mm)
Outer Dimensions:
Cuff Dimensions:
Accessories:
Approximately 5 1/2" x 19" (140 mm x 480 mm)
Arm cuff, carrying case, instruction manual
NOTE: These specifications are subject to change without notice.
Download from Www.Somanuals.co1m7. All Manuals Search And Download.
TITLE
For Customer Service
Call Toll Free
1-800-634-4350
Manufactured by OMRON
Distributed by:
OMRON HEALTHCARE, INC.
300 Lakeview Parkway
Vernon Hills, Illinois 60061
Copyright © 2000 OMRON HEALTHCARE, INC.
?000000-0X
18
|