Omron Healthcare Blood Pressure Monitor HEM 757 User Manual

R
INSTRUCTION MANUAL  
Automatic Digital Blood Pressure Monitor  
with IntelliSense™  
Model  
HEM-757  
DIGITAL BLOOD PRESSURE MONITOR  
WITH FUZZY LOGIC HEM-757  
SYSTOLIC  
DIASTOLIC  
READY TO  
MEASURE  
REPLACE  
PULSE  
/min  
BATTERIES  
PROGRESS  
ON/OFF  
START  
 
INTRODUCTION  
Thank you for purchasing the Omron HEM-757 IntelliSenseTM  
Automatic Blood Pressure Monitor. Although your blood pressure  
fluctuates  
throughout the day, no self-adjustment is necessary as the monitor  
intelligently adjusts to your changing conditions. The IntelliSenseTM  
monitor uses fuzzy logic intelligence to sense both your systolic  
and diastolic blood pressure values, and the result is personalized  
inflation with every reading. Arm discomfort and misreading due to  
improper cuff inflation are virtually eliminated.  
Your new OMRON monitoruses the oscillometric method of blood  
pressure measurement. This means the monitor detects your  
blood's movement through your brachial artery and converts the  
movements into a digital reading. At the push of a button, the  
monitor automatically fills the cuff with air and displays your  
systolic, diastolic, and pulse measurements on an extra large  
display panel.  
Please read this instruction manual thoroughly before using  
the Omron IntelliSenseTM Automatic Blood Pressure Monitor.  
For specific information on your own blood pressure,  
CONTACT YOUR PHYSICIAN.  
1
 
KNOW YOUR UNIT  
IntelliSense™ Progress Indicator  
Systolic Blood  
Pressure Display  
ON/OFF Button  
READY TO  
DIGITAL BLOOD PRESSURE MONITOR  
WITH FUZZY LOGIC HEM-757  
ON/OFF  
MEASURE  
REPLACE  
START Button  
BATTERIES  
PROGRESS  
SYSTOLIC  
START  
DIASTOLIC  
PULSE  
/min  
Memory Call Button  
Pulse Display  
Air Jack  
Diastolic Blood  
Pressure Display  
Air Plug  
(Connects to the air jack)  
Arm Cuff  
Air Tube  
Green Cuff Marker  
Carrying Case  
2
 
QUICK REFERENCE GUIDE  
1. Avoid eating, smoking, and exercising for at least 30 minutes  
before taking a measurement.  
2. Put your left arm through the cuff loop making sure that the  
bottom edge of the cuff is approximately one-half inch above the  
elbow.  
3. Pull the end of the cuff so that the entire cuff is evenly tightened  
around your arm and press the sewn hook material firmly against  
the pile side of the cuff. Make sure that your skin will not be  
pinched in the D-ring once you inflate the cuff bladder.  
Immediately deflate the cuff bladder and readjust the cuff if  
skin becomes pinched.  
4. Sit in a chair with your feet flat on the floor and place your left  
arm on a table so that the cuff is at the same level as your heart.  
5. Press the ON/OFF button.  
6. After the Heart Symbol ( ) appears next to a zero on the digital  
panel, press the START button.  
7. Remain still throughout the entire reading.  
8. When measurement is complete, the monitor alternately displays  
your blood pressure and pulse on the digital panel and  
automatically deflates the cuff.  
9. Wait 2-3 minutes before taking another blood pressure  
measurement. You may require more rest time between readings  
depending on your individual physiological characteristics.  
3
 
BATTERY INSTALLATION/REPLACEMENT  
1. Slide the battery cover off in  
the direction of the arrow.  
2. Install or replace 4 "AA" size  
batteries so + (positive) and  
– (negative) polarities match  
the polarities of the battery  
compartment as indicated.  
3. Replace the battery cover.  
4. If the Low Battery Indicator  
(
) appears on the display,  
Low Battery Indication  
replace all four batteries.  
Long-life alkaline batteries  
are recommended.  
5. Remove the batteries if the  
monitor will not be used for  
an extended period of time.  
4
 
SUGGESTIONS BEFORE BLOOD PRESSURE MEASUREMENT  
1. Avoid eating, smoking, and exercising for at least 30 minutes  
before taking a measurement. Also rest for at least 5 minutes  
before taking a reading.  
2. Stress raises blood pressure. Avoid taking measurements during  
stressful times.  
3. Remove tight-fitting clothing from your left upper arm.  
4. Measurement should be taken  
in a quiet place and you  
should be in a relaxed, seated  
position. Rest your left arm on  
the arm of a chair or on a table  
so that the cuff is at the same  
level as your heart.  
5. Remain still and do not talk during the measurement.  
6. Keep a record of your blood pressure and pulse measurements  
for your doctor. Remember, a single measurement does not  
provide an accurate indication of your true blood pressure. You  
need to take and record several measurements over a period of  
time. Try to measure your blood pressure at the same time each  
day for consistency. Blood pressure measurements fluctuate  
considerably.  
7. Wait 2-3 minutes between successive measurements. Waiting  
allows the engorged blood vessels to return to normal. You may  
require more rest time between readings depending on your  
individual physiological characteristics.  
5
 
HOW TO APPLY THE ARM CUFF  
1. Insert the air plug of the arm cuff  
into the air jack of the monitor.  
2. The cuff should be assembled  
correctly when it is removed from  
the box. If it is not, pass the end  
of the cuff furthest from the  
tubing through the metal D-ring  
to form a loop. The smooth cloth  
should be on the inside of the  
Pile Side Material  
Sewn Hook Material  
cuff loop.  
3. If the cuff is assembled correctly,  
the sewn hook material material  
will be on the outside of the cuff  
loop and the metal D-ring will not  
touch your skin.  
D-Ring  
4. Put your left arm through  
the cuff loop. The bottom  
of the cuff should be  
approximately 1/2"  
Bottom Edge 1/2 Inch  
above the elbow. The  
cuff tubing should be  
positioned along the  
middle of the inside of  
your arm as shown.  
Cuff Tube  
Cuff Tab  
5. Pull the cuff so that the top and  
bottom edges are tightened  
evenly around your arm.  
6
 
HOW TO APPLY THE ARM CUFF  
6. When the cuff is positioned  
(Left arm)  
correctly, press the sewn  
hook material FIRMLY  
against the pile side of the  
cuff.  
7. The cuff should be  
wrapped so that it fits  
snugly and stays in place.  
You should be able to fit  
one finger between the cuff  
and your arm.  
8. Sit in a chair with your feet  
flat on the floor and rest  
your arm on a table so that  
the cuff is at the same level  
as your heart.  
9. Relax your arm and turn  
your palm upward.  
10. Be sure there are no kinks  
in the air tubing.  
7
 
HOW TO TAKE A READING  
Please read "A Few Suggestions Before Blood Pressure Measurement" and  
"How To Apply The Arm Cuff" before taking a reading.  
1. Press the ON/OFF Button.  
a) All display symbols appear for  
approximately one second. This  
is the initial LCD test.  
b) The display symbols disappear  
and the Deflation Indication  
Initial LCD Test  
Symbol ( ) starts to flash.  
c) When the monitor completes the  
necessary preparations before  
measurement, the Heart Symbol  
(
) appears next to a zero.  
Deflation Indication  
NOTE: Wait for the Heart Symbol (  
before taking a measurement.  
)
2. Press the START button. As the  
cuff begins to inflate, the  
IntelliSense™ monitor  
Ready to Measure  
automatically determines your  
ideal inflation level. During this  
process, the IntelliSense™  
symbol ( ) will be flashing on  
the display. Because this monitor  
detects the pulse even during  
inflation, do not move your arm  
and remain still during the  
measurement.  
START  
Display of measurement process  
A
mark indicates the  
Inflation started.  
measurement process.  
When the Start Button is pushed,  
a
mark flushes.  
8
 
HOW TO TAKE A READING  
3. The IntelliSense™ Progress  
Indicator ( ) will flash on the  
display showing that the  
monitor’s fuzzy logic intelligence  
is determining your personal  
measurements.  
NOTE: Because blood pressure is being  
measured even during deflation, it is  
important to remain still until the reading  
is fully completed.  
Display of measurement process  
A
mark indicates the  
measurement process. During  
the inflation, a mark turns on.  
During the deflation, a mark  
turns on and one each mark  
turns off with the progress of  
approximately every 1/3 of the  
process.  
4. When the measurement is  
complete, a Heart Symbol (  
)
flashes and the arm cuff deflates.  
5. At the end of the measurement,  
your blood pressure and pulse  
are alternately displayed.  
6. Press the ON/OFF button to turn  
the monitor off.  
NOTE: If you forget to turn the monitor  
off, the monitor will automatically power  
off after five minutes.  
ON/OFF  
9
 
HOW TO TAKE A READING  
Special instructions regarding the inflation of  
the IntelliSense™ Monitor:  
1. In rare circumstances the user’s  
physiological characteristics may  
require a higher level of inflation.  
These circumstances may  
include weak pulse or an  
unusually high systolic blood  
pressure. When this occurs, the  
monitor may reinflate to a value  
of about 30 mmHg higher than  
the previous inflation value.  
Reinflation occurs only once.  
2. If your systolic pressure is known  
to be more than 220 mmHg,  
push and hold the START button  
START  
until the monitor inflates 30 to 40  
mmHg higher than your  
suspected systolic pressure.  
Continue to press button  
NOTE: Do not apply more  
pressure than  
necessary.  
3. If you want to stop the  
measurement, push the ON/OFF  
button. The monitor will stop  
inflating and start deflating  
rapidly, then the monitor will  
power off.  
10  
 
HOW TO USE THE MEMORY FUNCTION  
How to Use the Memory Function  
This monitor has the function to store fourteen measurement values.  
Push the  
Button while a ( ) mark is displayed.  
• Maximum of fourteen stored measurement values can be called  
out.  
• For the fifteenth measurements and after, the values from the  
oldest measurement will be deleted.  
Previous value  
Value before  
the previous  
measurement  
Value of  
fourteen  
measurements  
before  
If you want to delete all the stored data  
You can delete all the stored data while the ( ) mark is displayed.  
START  
Push these two buttons at  
the same time.  
Display of memory deletion  
Download from Www.Somanuals.co1m1. All Manuals Search And Download.  
HOW TO USE THE OPTIONAL AC ADAPTER  
(ACCESSORY ITEM SOLD SEPARATELY)  
1. When using AC power, use only the exclusive AC adapter  
(Model HEM-ADPT1) which can be purchased from:  
Omron Healthcare, Inc.  
300 Lakeview Parkway  
Vernon Hills, IL 60061  
Toll Free Phone #: 1-800-634-4350  
Plug  
2. Insert the AC adapter plug into  
the jack on the side of the  
monitor.  
3. Insert the AC adapter plug into  
a 120V AC outlet (50 - 60  
Jack  
cycles).  
Option AC Adapter  
4. To remove the AC adapter, disconnect the adapter plug from the  
AC outlet first and then disconnect the cord from the monitor’s  
jack. To avoid possible damage to the monitor, use only the  
exclusive AC adapter (Model HEM-ADPT1) available from  
Omron Healthcare, Inc. Other adapters may vary in output  
voltage and polarities.  
AC Adapter Output  
6 V, 500 mA  
NOTE: The monitor is designed not to draw power from the  
batteries when the AC adapter is in use.  
12  
 
TROUBLESHOOTING  
Error Indicator  
Cause  
Correction  
Wait 10 minutes before taking  
another measurement. Repeat  
steps listed under "How to Take a  
Reading." If "EE" is displayed again,  
take another measurement by  
pressing and holding the START  
button as listed in step 2 under  
"How to Take a Reading."  
Cuff under-inflated.  
Movement during  
measurement.  
Carefully read and repeat steps  
listed under "How to Take a  
Reading."  
Cuff over-inflated (more than  
300 mmHg).  
Battery voltage is excessively Replace all four "AA" batteries with  
low.  
new batteries.  
Please contact the store where you  
have purchased this unit or the  
nearest OMRON dealer.  
Trouble caused by abnormal  
memory function.  
If you continue to have difficulty obtaining a reading, call  
OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1-800-634-4350.  
Please tell the representative you have an OMRON Model HEM-757.  
Download from Www.Somanuals.co1m3. All Manuals Search And Download.  
CAUTION  
CAUTION:  
Changes or modifications not expressly approved by Omron Healthcare, Inc.  
could void the user’s authority to operate this product.  
NOTE:  
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)  
This product has been tested and found to comply with the limits for a Class B  
digital device, pursuant to part 15 of the FCC rules.  
These limits are designed to provide reasonable protection against harmful  
interference in a residential installation. The product generates, uses, and can  
radiate radio frequency energy and, if not installed and used in accordance  
with the instructions, may cause harmful interference to radio communications.  
However, there is no guarantee that interference will not occur in a particular  
installation. If the product does cause harmful interference to radio or television  
reception, which can be determined by turning the product on and off, the user  
is encouraged to try to correct the interference by one or more of the following  
measures:  
• Reorient or relocate the receiving antenna.  
• Increase the separation between the product and receiver.  
• Connect the product into an outlet on a circuit different from that to  
which the receiver is connected.  
• Consult the dealer or an experienced radio/TV technician for help.  
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)  
This digital apparatus does not exceed the Class B limits for radio noise  
emissions from digital apparatus as set out in the interference-causing  
equipment standard entitled "Digital Apparatus", ICES-003 of the Canadian  
Department of Communications.  
Cet appareil numérique respecte les limites de bruits radioeléctriques  
applicables aux appareils numériques de Classe B prescrites dans la norme  
sur le matériel brouilleur: "Appareils Numériques", NMB-003 èdictée par le  
ministre des communications.  
14  
 
CARE AND MAINTENANCE  
To protect your monitor from damage, please AVOID the  
following:  
• Subjecting your monitor to extreme temperatures, humidity,  
and direct sunlight.  
• Folding the cuff and tubing tightly.  
• Inflating the monitor over 280 mmHg.  
• Disassembling the monitor.  
• Subjecting the monitor to strong shocks (for example,  
dropping the monitor on the floor.)  
• Do not clean the monitor with volatile liquids.  
THE MONITOR SHOULD BE CLEANED WITH A SOFT,  
DRY CLOTH.  
Download from Www.Somanuals.co1m5. All Manuals Search And Download.  
FIVE YEAR LIMITED WARRANTY  
Your HEM-757 IntelliSense™ Automatic Blood Pressure Monitor is  
warranted to be free from manufacturing defects for a period of five  
years under normal use. The five year warranty excludes the  
monitor cuff. The cuff is warranted for a one year period. This  
warranty extends only to the original retail purchaser.  
Should repair be needed within the warranty period, ship the unit  
prepaid to Omron Healthcare, Inc., 300 Lakeview Parkway,  
Vernon Hills, IL 60061, Attn: Service Dept., together with $5.00  
for return shipping and insurance. Be sure to include the model  
number of your unit and your phone number on any  
correspondence.  
We will either repair or replace (at our option) free of charge any  
parts necessary to correct defects in the materials or workmanship.  
The above warranty is complete and exclusive. The warrantor  
expressly disclaims liability for incidental, special, or consequential  
damages of any nature. (Some states do not allow the exclusion or  
limitation of incidental or consequential damages, so the above  
warranty may not apply to you.)  
Any implied warranties arising by the operation of law shall be  
limited in duration to the term of this warranty. (Some states do not  
allow limitations on how long an implied warranty lasts, so the  
above limitation may not apply to you.)  
This warranty gives you specific legal rights, and you may have  
other rights which vary from state to state. As a condition to  
operation of your warranty, the enclosed registration card must be  
completed and sent to us within 10 days from the date of purchase.  
FOR CUSTOMER SERVICE CALL TOLL FREE:  
1-800-634-4350  
16  
 
SPECIFICATIONS  
Model:  
HEM-757  
Display:  
LCD Digital Display  
Measurement Range:  
Pressure: 0 to 280 mmHg  
Pulse: 40 to 180/minute  
Pressure: ±3 mmHg or 2% of reading  
Pulse: ±5% of reading  
Automatic by electric pump  
Automatic deflation system  
Accuracy/Calibration:  
Inflation:  
Deflation:  
Rapid Pressure Release:  
Pressure Detection:  
Measurement Method:  
Pulse Wave Detection:  
Power Source:  
Active electromagnetic control valve  
Electrostatic capacity type pressure sensor  
Oscillometric method  
Capacitive pressure sensor  
4 "AA" batteries or AC adapter (sold separately)  
Approximately 300 uses  
Battery Life:  
Operating Temperature/Humidity: 50°F to 104°F (10°C to 40°C)  
30 to 85% RH maximum  
Storage Temperature/Humidity:  
–4°F to 140°F (–20°C to 60°C)  
10 to 95% RH maximum  
Console Weight:  
Approximately 19 oz (530 g) not including batteries  
Approximately 4 1/2" (l) x 7" (w) x 2 5/6" (h)  
(115 mm x 177 mm x 72 mm)  
Outer Dimensions:  
Cuff Dimensions:  
Accessories:  
Approximately 5 1/2" x 19" (140 mm x 480 mm)  
Arm cuff, carrying case, instruction manual  
NOTE: These specifications are subject to change without notice.  
Download from Www.Somanuals.co1m7. All Manuals Search And Download.  
TITLE  
For Customer Service  
Call Toll Free  
1-800-634-4350  
Manufactured by OMRON  
Distributed by:  
OMRON HEALTHCARE, INC.  
300 Lakeview Parkway  
Vernon Hills, Illinois 60061  
Copyright © 2000 OMRON HEALTHCARE, INC.  
?000000-0X  
18  
 

Miele Vacuum Cleaner S 300I User Manual
Milwaukee Drill 2432 20 User Manual
National Instruments Network Card Xmath User Manual
NEC Computer Monitor MD213MC User Manual
Nilfisk Advance America Lawn Sweeper SR 1700D 2W D User Manual
Nokia Satellite TV System 9650S User Manual
Nortel Networks Telephone 4145Ex User Manual
O2 Innovations Network Router DSL Router User Manual
Oregon Scientific Marine Radio BAR628HG BAR628HGU User Manual
Panasonic Answering Machine KX TG5838AL User Manual